Persona IQ Smart Knee

Zimmer Biomet / Canary Medical

Not determinedIndexed from FDA records

MR Conditional — conditions not yet documented here. We have a Conditional label for this device but the specific scan conditions (field strength, SAR / B1+rms, monitoring requirements) aren't in our records yet. Treat as Not determined and let your MRI team confirm before any scan.

Here's what to do next — you're going to be okay:

  1. Bring your device card (the wallet card from your surgery) to your appointment. Your MRI team can identify any device from it.
  2. Call the manufacturer's patient line — the number is on your card. They answer device questions every day.
  3. Tell your imaging center about your device when you book. They screen every patient before every scan — no one will scan you unless it's safe.

Don't schedule an MRI based on this page alone — your MRI team verifies every device.

What this means for you

MR Conditional — but the conditions are not yet on file. The label exists. The scan rules (field, SAR, monitoring) do not. Treat as Not determined and let your MRI team confirm before any scan.

Here's what to do next — you're going to be okay:

  1. Bring your device card (the wallet card from your surgery) to your appointment. Your MRI team can identify any device from it.
  2. Call the manufacturer's patient line — the number is on your card. They answer device questions every day.
  3. Tell your imaging center about your device when you book. They screen every patient before every scan — no one will scan you unless it's safe.

For information only — your MRI team makes the final decision before any scan. They check every patient, every time.

This recall has been resolved
⚠️

FDA Recall Alert

Class I — Most Serious

The drill bit used to drill the hex pre-drilled hole in the screw head of the Tibial Augment Bolts included in the Vanguard 360 Revision Knee System was larger than allowable per the process specification. As a result, the hex geometry can still be visible, but there are radial cuts into the hex flat. Therefore, when the bolt is inserted into the augment, the screwdriver is unable to catch on the hex geometry, or the screwdriver strips the remaining hex geometry, and the bolt is unable to be fully tightened. The bolt stripping or the bolt not being fully tightened will be noticed by the surgeon intraoperatively when he or she is unable to screw the augment bolts into the augment. This possible event can result in an intraoperative delay resulting in increased time under anesthesia.

💡 What this means for you

This is the most serious type of recall. The FDA determined this product could cause serious health consequences. Talk to your doctor about what this means for you.

8/7/2016 · View on FDA.gov

This recall has been resolved
⚠️

FDA Recall Alert

Class I — Most Serious

Contents of the package do not match the product labeling. Product is labeled: Vanguard Distal Femoral Augment 75 X 5 RL/LM Carton contains: Vanguard Distal Femoral Augment 75 X 5 LL/RM Risks: Adverse health outcomes that could occur as a result of this event are a delay less than 30 minutes and a different size augment being used to complete the procedure. As the augment will not assemble with femoral component due to its shape, it would be easily identified during surgery.

💡 What this means for you

This is the most serious type of recall. The FDA determined this product could cause serious health consequences. Talk to your doctor about what this means for you.

8/8/2016 · View on FDA.gov

⚕️ This information is for reference only. A recall does not necessarily mean your device needs removal. Consult your healthcare provider for medical advice.

Regulatory Info

510(k) NumberK251834 ↗
Clearance Date8/14/2025
Product CodeHSX
All Recalls Resolved

2 past recalls — all corrective actions completed

Data sourced from openFDA. This is for reference only — consult your healthcare provider for medical advice.

Category

Joint Replacement

Body Region

Knee

MRI Scan Parameters

Max Field Strength3.0T

MRI Conditions

MR Conditional at 1.5T and 3.0T.

Indexed from FDA records. Conditions on this page have not been verified against the manufacturer's instructions for use.